|
|
JEOL 4000EX High Resolution Electron Microscope (HREM) 400kV
Alignment Procedure
Check that the objective and selected area apertures are out.
Check the vacuum: IGP1: 5.4 x 10-5 black scale (max 7 x 10–5)
IGP2: 0.8 x 10-5 black scale (max 3 x 10–5)
Check that the green READY light is lit and BRIT TILT button is lit.
- HT should be on at 350 KV; turn on at 300 KV, if not on, and step up to 350 in 1 KV steps.
- Step up KV to 400, 1 KV at a time.
- Open AIR LOCK. Wait for green filament light.
- Increase Filament temperature SLOWLY, let it sit @ 4.5 for 20 min, then increase the BEAM CURRENT 0.1
mA at a time. Saturate the filament, usually found at the silver anchor; it may vary. Filament image should be visible and symmetrical. Adjust with GUN DEFs if not symmetrical.
- Slowly converge the beam with BRIGHTNESS, and center with SHIFTS.
- At spot size 1 and 25KX, center the converged beam with GUN SHIFT.
- Increase the spot size to 5, converge the beam and center with SHIFT.
- Repeat centering at spot size 1 (GUN SHIFT), then spot size 5 (SHIFT), to center gun with condenser lens.
- Choose spot size 2.
- Correct astigmatism with the COND STIGMATOR DEFLECTOR.
- Turn down the filament; close AIR LOCK.
- Check that specimen drives are at 0,0. Insert sample into column:
- Exchange mechanism lever on O, for open.
- DOWN, to specimen holder.
- GRAB (unlabelled knob)
- UP
- IN
- DOWN
- RELEASE (unlabelled knob)
- UP
- OUT
- Exchange mechanism lever to P, for pump.
- Open AIR LOCK, increase filament temperature, SLOWLY.
- Set OBJ lens current to focus (@400 KV = 6.65)
- Adjust z-axis for minimum contrast.
- Increase mag to 200K X and correct for HT wobble (BRIT TILT DEF). If wobble is gross, start at a lower mag.
- In the DIFF mode, insert the objective aperture, center it around the illumination.
- At 200K X or more, correct for objective astigmatism (OBJ STIG DEF).
SHUT DOWN
- Turn off the FILAMENT, slowly.
- Close the AIR LOCK.
- Check that specimen drives are at 0,0. Remove the sample from the column:
- Exchange mechanism lever on O, for open.
- Wait for green SPEC light.
- IN
- DOWN
- GRAB (unlabelled knob)
- UP
- OUT
- DOWN, to specimen holder.
- RELEASE (unlabelled knob)
- UP
- Exchange mechanism lever to P, for pump.
- Leave HT on and KV at 350.
- Log in the state of the microscope as it was left.
TO REMOVE FILM:
Wear gloves !
- Retrieve the empty plate holder box from the desiccator or the darkroom; a dark bag should be with it, to transport the exposed negatives across the lighted hallway.
- Turn on dry nitrogen at the tank.
- Turn the handle on the front of the camera door clockwise 90o.
- Turn out the room light.
- Wait for the camera door to pop open.
- Pull out the drawer containing the film boxes, by the handle.
- The first box contains exposed film (to be developed), remove it from the drawer, put the box in the dark bag.
- If the FILM box has less than 10 pieces of film, replace it with a fresh box from the desiccator.
- Close the camera door securely: it will require a strong push.
- Turn the handle on the door counterclockwise 90 o; the vacuum system should be audibly pulling on the camera chamber.
- Turn off the dry nitrogen tank.
- Take both boxes to the darkroom: either refill the FILM box with fresh film, or tell someone who knows how to fill it that it needs to be done.
Return to CNC
|